Seal Check Test for Sealing Quality Testing of All Sealers for H2O2/Plasma A170 GM170SCT– SHORT CODE: A170

It is a test that the devices work in accordance with EN-ISO 11607-2 standard and factory settings.



222-110-070_A170 GM170SCT SEAL-CHECK-Tyvek® (100 pcs in 1 Package

  • These tests are designed as two models and should be selected according to the type of pouch you use.
  • These tests should be performed daily before each use, before the sealer is used. These tests are defined in EN-ISO 11607-2 and DIN 58953.
  • Take one Seal-Check sheet out of its package and place it in a pouch with a width of at least 20 cm. Give it to your device together with the pouch. Successful or, if any, incorrectly sealed areas will be clearly seen in the pouch that comes out of the process. For the maintenance of your device in case of malfunction, contact the place where you bought your device and have the parts replaced to return your device to factory settings.
  • A successful test result shows that your device has been sealed in accordance with EN ISO 11607-2 standard. The test result can be documented on a daily basis and serves as evidence in audits.

Under the "Related Products" heading below, you can see the accessories that can be used with this product, for detailed technical data; you can download the brochure, technical specifications (Membership may require) and quality certificates, etc. of that product from the "Files" tab above. You can watch the operation of the product from the "Videos" tab, if available.

Related products